Sierra Monitor Corporation Carbon Monoxide Alarm 2050 10 User Manual

GAS MONITORS  
MODELS 2050-00 & 2050-10  
MODELS 2060-00 & 2060-10  
 
TABLE OF CONTENTS  
SECTION  
PAGE  
1.PRODUCT DESCRIPTION  
1.1 Introduction  
1
1
1
1
2
2
2
1.2Application  
1.3 Configuration  
1.4 Temperature Range  
1.5 Sensor  
1.6 RemoteAlarm  
2. QUICK START  
2.1 Overview  
2.2 Wiring  
3
3
3
3. OPERATION  
3.1 Introduction  
3.2Alarms  
4
4
4
4
3.3 Outputs  
4. ALARM CALIBRATION  
5
5
5
5
6
6
4.1 FactoryAlarm Calibration  
4.2 Frequency ofAlarm Calibration Check  
4.3 Alarm Calibration Check Process  
4.4 Equipment Required  
4.5 Alarm Calibration Procedure  
5. SERVICE  
5.1 Sensor Replacement  
7
6. INSTALLATION  
8
8
8
8
8
8
6.1 Gas Sensor Module Locations  
6.2 Mounting  
6.3 Wiring  
6.4 Power Supply  
6.5Alarm Devices  
7. ALARM CONFIGURATION  
8. SPECIFICATIONS  
10  
11  
11  
11  
9. ACCESSORIES AND REPLACEMENT PARTS  
10. LIMITED WARRANTY  
Table of Contents  
 
Instruction Manual  
1.0 PRODUCT DESCRIPTION  
1.1 Introduction  
Model  
2050-00  
2060-00  
Gas  
Ammonia  
Halocarbons  
TTL  
X
X
Relay  
X
X
The Model 2050 Ammonia and Model 2060 Refrigerant/  
Halocarbon Gas Monitors are continuous duty, low cost  
stationary monitors for detection of high levels (50-500  
PPM) of ammonia and halocarbon ("freons") refrigerant  
gases. See the specifications page for a full list of avail-  
able applicable halocarbons  
Table 1.1 Gas Monitor Configurations  
This manual provides instructions for both the 2050 and  
2060 series Gas Monitors, including the relay versions.  
factory setpoint for R12 is 500 ppm while the setpoint  
for R134 a is 50 ppm. See the specifications for more  
details.  
1.2 Application  
The Model 2050 / 2060 is cross sensitive to other com-  
bustible gases. If combustible gasses in excess of 50  
ppm are expected in the area please contact Sierra Moni-  
tor to determine amount of cross-sensitivity expected  
for the specific combustible gas. It is not recommended  
to use this unit in areas with combustible gases present.  
These monitors are primarily used in and around freez-  
ers, refrigerators, and compressor rooms to warn of leaks  
and dangerous levels of these gases.  
1.3 Configuration  
An alarm is activated when the concentration of gas  
exceeds the user-adjustable level. An alarm condition  
activates the alarm light on the front of the monitor and  
generates an output signal. On Models 2050-00 and the  
2060-00 the output signal a TTL (Transistor-Transis-  
tor-Logic signal) can activate a remote alarm through  
one of Sierra Monitor’sAlarm Panels. (The Model 2102  
Dual Channel Alarm Panel can be connected up to 500  
feet from the monitors. ) The models can be config-  
ured with a 0.5 amp, dry contact relay.  
The monitors are housed in a  
polystyrene enclosure that is suit-  
able for installation in freezer, re-  
frigerator or other applications.  
Test connections and adjustment  
controls are located on the out-  
side of the enclosure and are eas-  
ily used, even when the operator  
is wearing hand protection.  
The Model 2050 Ammonia moni-  
tor has a factory alarm setpoint of  
100 ppm NH3. This setpoint is  
user adjustible in the field over a  
range of 50 to 500 ppm NH3. The  
Model 2050 calibration can be  
checked with Hydrogen gas.  
The Model 2060 Halocarbon  
monitor can be calibrated with  
carbon monoxide and the factory  
setpoint varies for the specific  
type of halocarbon present. The  
Figure 1.1 Outline View - 2050/2060 Series  
2050/2060 Gas Monitors (5/04)  
Page:  
1
 
Instruction Manual  
1.4 Temperature Range  
The Model 2050 and 2060 have temperature compen-  
sated circuitry to provide stabilization in temperatures  
ranging from -4o to 158oF.  
However, for optimum operation the unit should be cali-  
brated at the normal temperature expected for the sensor  
location.  
1.5 Semiconductor-Type Sensor  
A solid-state semiconductor-type sensor and associ-  
ated electronic circuitry ensure trouble-free, long-term  
operation. All the electronic circuitry needed to operate  
the monitor, except the DC input power, is contained in a  
compact unit. There are no pumps, filters, or chemical  
cells to replace or maintain. Except for periodic calibra-  
tion to verify the alarm setting, no attention is required  
after installation. The user may adjust the alarm level by  
applying a different calibration gas concentration.  
A sensor self-check feature will flash the alarms on and  
off alternatively should the sensor fail (open circuit).  
The alarm flash includes LEDs, audible alarm, relay or  
alarm signal as applicable.  
1.6. RemoteAlarmAvailable  
If a remote alarm is desired, Sierra Monitor’sAlarm Pan-  
els, the Model 2102 Dual Channel Alarm Panels can be  
connected up to 500 feet away from the monitors.  
2050/2060 Gas Monitors (5/04)  
Page:  
2
 
Instruction Manual  
2.0 QUICK START  
2.1 Overview  
The Gas Sensor Module has been supplied factory  
calibrated and ready for immediate installation and  
operation. An installer familiar with installation and  
operation of gas detection products can use this  
section to begin immediate use of the monitor.  
During warm-up the monitor will, first, cycle  
through safe/alarm/safe condition at one hertz.  
This will be followed by a short period of continu-  
ous alarm before warm-up is completed.  
2.2.4 Configuration  
2.2 Wiring  
The default configuration for each module is to  
operate with a buzzer and a normally operating  
open (NOO) relay. The user can change this con-  
figuration using the jumpers provided. Refer to  
table 6.2.  
Each module requires four-conductor wiring (two  
wires for power and two wires for the signal).  
2.2.1. Model 2050-2060 Series  
The Model 2050-2060 series are designed to  
mount on any indoor vertical surface. Mount  
the monitor in the desired location.  
-
The installation must meet any hazard-  
ous environmental codes for AC/DC elec-  
trical instrumentation.  
2.2.2 Wiring Connection  
Terminal positions on the electronics board are  
as follows:  
Terminal  
TB1-1  
Function  
+ VDC (9 – 24)  
TB1-2  
GND (0VDC)  
TB2-1  
Relay NC (Normally Closed) or  
Output to Model 2102 Safe  
Relay Common  
Relay NO (Normally Open) or  
Output to Model 2102 Alarm  
Table 2.1  
TB2-2  
TB2-3  
2.2.3 Start-up & Operation  
To begin operation of the Gas Sensor Module  
provide 9 - 24 VDC from a regulated power  
supply such as one of the Sierra MonitorAlarm  
Panels, 2102-XX. Each time the sensor mod-  
ule is powered up it will perform a warm-up for  
2 - 60 minutes.  
Page:  
2050/2060 Gas Monitors (5/04)  
3
 
Instruction Manual  
3.0 OPERATION  
3.1 Introduction  
Trouble Alarm  
Under normal conditions the gas sensor module does  
not require operator or technician intervention. The  
following are conditions under which the module re-  
quires attention:  
Interrupted contact closure and “red/green” vi-  
sual indication (once the warm-up time is com-  
pleted) after unit has been in operation, indicat-  
ing either a failed sensor or calibration problems.  
- Routine periodic calibration  
- Sensor replacement on a planned schedule  
or when a sensor failure occurs.  
Buzzer  
Relay  
Relay  
Fail Safe  
- Periodic cleaning as necessary.  
On/Off  
Oscilating  
Oscilating  
3.2 Alarms  
Three alarm conditions are possible. These can be  
detected visually at the optional Sierra Monitor con-  
troller and at the module.  
3.3 Outputs  
For the non-relay configurations a 5 volt DC TTL  
signal is provided to activate remote alarms such  
as the Sierra MonitorAlarm Panels. Relay configu-  
ration alarm conditions described above are trans-  
mitted to the annunciator as relay contact closures  
from the monitor.  
Warm-Up Alarm  
Oscillating (On/Off) contact closure and “red/  
green” visual indication when power is first con-  
nected to the module, followed by a continuous  
closure of approximately 2-60 minutes indicat-  
ing warm-up.  
The alarm outputs are non-latching closed contacts.  
As soon as the gas concentration falls below the  
alarm set-point, the alarm will stop and the alarm  
output will cause the relay to de-energize and the  
contacts will open.  
Buzzer  
Relay  
Relay  
Fail Safe  
During warm-up and failed sensor alarms the re-  
lays contact and the display LED's oscillate at ap-  
proximately one hertz.  
On/Off  
Oscilating  
Oscilating  
Gas Alarm  
Sustained contact closure and solid red light  
(once the warm-up time is completed) indicat-  
ing the presence of gas at, or above, the pre-set  
alarm limit.  
Table 2.1  
Buzzer  
Relay  
Relay  
Fail Safe  
On  
Closed  
Open  
=
2050/2060 Gas Monitors (5/04)  
Page:  
4
 
Instruction Manual  
4.0ALARM CALIBRATION  
of each gas sensor module be verified monthly dur-  
ing the first three months of operation and then quar-  
terly. More frequent checks are necessary during  
periods of extreme humidity and temperature  
changes. The monitor should have operated con-  
tinuously (uninterrupted) for at least 24 hours prior  
to calibration adjustment.  
4.1 FactoryAlarm Calibration  
The module has been factory calibrated to alarm as  
indicated in Table 4.1 and as marked on the cali-  
bration label shipped with the module.  
Model  
Time  
Gas  
cc/min  
4.3AlarmCalibrationCheckProcess  
2050-XX  
2060-XX  
100sec. 100 ppm NH3  
50  
The output signal of the gas sensor module is calibrated  
using a calibration gas mixture containing a known  
concentration of the gas of interest and a balance of air,  
or using a substitute gas. The concentration of the span  
gas must be equal to the alarm point desired.  
60 sec. 500 ppm R11 50  
500 ppm R12  
50 ppm R22  
50  
50  
100 ppm R113 50  
70 ppm R123 50  
50 ppm R134a 50  
40 ppm R141b 50  
40 ppm R142b 50  
The equivalent substitute gas is 80 ppm Hydrogen (H2)  
for the Model 2050 and 150 ppm Carbon Monoxide  
(CO) for the Model 2060.  
40 ppm R500  
70 ppm R502  
50  
50  
Calibration requires application of the span gas to the  
sensor and adjustment of the sensitivity adjustment po-  
tentiometer.  
Table 4.1  
Warning: During calibration the alarm will turn on and  
remote alarms connected to the alarm relays will be  
activated. Disable the remote alarm if necessary.  
4.2 Frequency of Alarm Calibration Check  
The manufacturer recommends that the calibration  
Fig. 4-1 Location of Calibration Voltmeter Connections and Adjustments  
Page:  
2050/2060 Gas Monitors (5/04)  
5
 
Instruction Manual  
4.4EquipmentRequired  
The following tools and equipment will be required for  
calibration:  
-JewelersScrewdriver  
- Calibration Gas  
- Calibration gas delivery system  
For accurate calibration use a gas mixture at the re-  
quired concentration mixed in an air balance, rather  
than with an inert gas like nitrogen. This gas and the  
required delivery equipment such as the Model 1200-  
26 Calibrator is available from Sierra Monitor Corpo-  
ration.  
4.5Alarm Calibration Procedure  
1. The monitor should be in the safe condition prior  
to calibration (green LED “ON”).  
-
Be sure that the area is non-hazardous be-  
foreproceeding.  
2. With the monitor in operation, expose the sensor  
to a sample of calibration gas. Apply calibration  
gas directly to the sensor insuring that the delivery  
cup loosely covers the sensor. Use 80 PPM H2 for  
the Model 2050 and 150 PPM CO for the Model  
2060.  
3. The red LED on the face of the monitor should turn  
"on". If the alarm occurs within 18-20 seconds and  
stops within one minute of the removal of the cali-  
bration gas, the monitor is in calibration and re-  
quires no further adjustment.  
4. If the monitor fails to alarm within 90 seconds of  
the application of the gas adjust the potentiometer  
counter clockwise until the alarm turns "on".  
5. If the monitor fails to stop alarming within one  
minute of the removal of the gas adjust the poten-  
tiometer clockwise until the alarm stops.  
6. After adjustment, repeat steps 4 and 5 to verify cali-  
bration.  
7. When the calibration is complete, reconnect any  
alarm equipment as necessary.  
2050/2060 Gas Monitors (5/04)  
Page:  
6
 
Instruction Manual  
5.0 SERVICE  
5.1 Sensor Replacement  
The gas sensor needs replacement when:  
2. Remove the cover from the main enclosure.  
3. Unplug the sensor connector on the lower  
right hand corner of the electronics board.  
-
-
-
It is no longer possible to obtain correct  
calibration  
4. Unscrew the sensor assembly from the end of the  
enclosure.  
The failed sensor alarm (oscillating  
red/green LED) is on  
5. Reverse the preceding steps to install the new sen-  
sorassembly.  
The sensor output signal is noisy,  
causingincorrectgasalarms.  
6. Restore power and allow a minimum of 24 hours  
for stabilization before recalibration. The calibra-  
tion should be re-verified after 6 days of continu-  
ous operation.  
To replace the sensor:  
1. Confirm that the system power has been  
removed.  
Figure 5-1 Model 2050/2060 Electronics Board Connections  
Page:  
2050/2060 Gas Monitors (5/04)  
7
 
Instruction Manual  
6.0 INSTALLATION  
6.1 Gas Sensor Module Locations  
6.4 Power Supply  
The gas sensor module is a diffusion type sensor  
that should be located close to the anticipated source  
or destination of the gas hazard. For heavy gases  
such as H2S install the module within 24 inches of  
the ground. For lighter gases such as CO and com-  
bustible gases use a higher elevation.  
The power supplied by the controlling device or an  
external power supply must meet the following speci-  
fications:  
Voltage:9 - 24 VDC  
Current: 250 mA  
After optimum locations are determined based on 6.5Alarm Devices  
the above recommendations, consideration should  
be given to placing the sensors in locations that are  
accessible for calibration service. Slight adjustments  
to the location of the sensor may have little impact  
on effectivity but major effect on accessibility.  
The internal relay is rated at 0.5Amp, normally open  
dry contact rated at 100 VDC/130 VAC.  
6.2 Mounting  
To mount the monitor, remove the four svrews in the  
cover plate and move the cover plate aside. Use  
screws through the monitoring holes visible in the  
four corners of the housing to mount the monitor to  
a wall.  
- The installation must meed any hazardous envi-  
ronmental codes forAC/DC electrical instrumenta-  
tion.  
6.3 Wiring  
Interconnect wiring from the controller to the module  
is by 4 conductor 22 AWG (or lower AWG) cable,  
conduit as necessary. Shielding is not required.  
For installations where the distance from the con-  
troller to the sensor is greater than 500 feet, 18  
AWG cable is recommended.  
Terminal  
Function  
TB1-1  
TB1-2  
TB2-1  
+ VDC (9 – 24)  
GND (0VDC)  
Relay NC (Normally Closed) or  
Output to Model 2102 Safe  
Relay Common  
Relay NO (Normally Open) or  
Output to Model 2102 Alarm  
Table 6.1  
TB2-2  
TB2-3  
The terminal strip on the electronics board in the  
module. The wiring must be connected as indi-  
cated in Figure 6.1 depending upon the controller  
or relay configuration being used.  
2050/2060 Gas Monitors (5/04)  
Page:  
8
 
Instruction Manual  
Figure 6.1  
Interconnect Wiring Table  
Page:  
9
2050/2060 Gas Monitors (5/04)  
 
Instruction Manual  
7.0 ALARM CONFIGURATION  
Model 20X or 200X Configured with Alarm Relay Output  
Configuration Jumpers  
Relay Operation  
Normally  
Buzzer Active  
Yes No  
Default  
Install Jumper Remove Jumper  
Normally  
Energized  
Not Energized  
Default  
Install Jumper  
JP2 pins 1-2  
Install Jumper  
JP2 pins 2-3  
JP6  
JP6  
(Best for 20X)  
Wiring Terminations  
Power Supply  
Output Terminals  
TB1-1  
+ DCV  
TB1-2  
0 DC  
TB2-1  
NC  
TB2-2  
Common  
TB2-3  
NO  
(9-24 VDC)  
Common  
Model 20X or 200X Configured for Interface to Model 2102 Alarm Panel  
Configuration Jumpers  
Buzzer Active  
Yes No  
Default  
Install Jumper Remove Jumper  
Relay Operation  
Interface Selection  
Normally  
Normally  
Model 2102  
Not Energized  
Default  
Energized  
N/A  
N/A  
Install Jumpers  
JP3 pins 1-2  
JP2 pins 1-2  
JP6  
JP6  
(Best for 20X)  
Wiring Terminations  
Output Terminals  
Power Supply  
TB1-1  
+ DCV  
TB1-2  
0 DC  
TB2-1  
Safe  
TB2-3  
Alarm  
TB2-2  
To Model 2102 To Model 2102 To Model 2102  
To Model 2102  
Not Used  
Terminal J5 -2 Terminal J5 -3 Terminal J5 -1 Terminal J5 -4 or  
or J6-2 or J6-3 or J6-1 J6-4  
Table 6.2  
Jumper Configuration and Wiring Terminations  
Relay Operation  
Normally Not Energized  
TB2-1 NC  
Connected  
Oscillating  
Connected  
TB2-2 Common  
TB2-3 NO  
No Power  
Sensor Fail  
Safe  
Open  
Open  
Alarm  
Open  
Connected  
Normally Energized - Fail Safe  
TB2-1 NC  
Open  
TB2-2 Common  
Connected  
TB2-3 NO  
Open  
No Power  
Sensor Fail  
Safe  
Oscillating  
Connected  
Alarm  
Connected  
Open  
Table 6.3  
Relay Operating States  
2050/2060 Gas Monitors (5/04)  
Page:  
10  
 
Instruction Manual  
8.0 SPECIFICATIONS  
Calibration Gas:  
2050 Series  
Indicators:  
Green LED for Monitor “ON”/Safe  
Red LED for Alarm  
80 ppm Hydrogen (H2)  
150 ppm Carbon Monoxide (CO)  
2060 Series  
Sensor Type:  
Solid-state metal oxide semiconduc-  
tor  
Power:  
9 VDC, 250 mA  
Signal Output::  
2050-00/2060-00  
Alarm Setpoint::  
2050 Series  
Nominal 5 volt DC, source 25 mA.  
SAFE and ALARM Signals  
0.5 AMP dry contact, 100 VDC, 130  
VAC  
100 ppm NH3  
2060 Series  
R11  
R12  
R22  
R113  
R123  
R134a  
R141b  
R142b  
R500  
R502  
500 ppm  
500 ppm  
50 ppm  
100 ppm  
70 ppm  
50 ppm  
40 ppm  
40 ppm  
40 ppm  
70 ppm  
Response Time:  
Less than 90 seconds  
Operating Temperature Range:  
-4 to 158 F (-20 to 70 C)  
Enclosure Material: Polystyrene  
o
o
o
o
Size:  
5.6 x 3.1 x 2.1 in.  
(14.2 x 7.9 x 5.3 cm)  
0.3 lb (0.15 Kg)  
Weight:  
User adjustable with calibration over  
range of 50 to 500 ppm  
9.0 ACCESSORIES AND REPLACEMENT PARTS  
Alarm Panels  
Replacement parts  
Model 2102-00 Alarm Panel - 2 channel  
Model 2102-01 Alarm Panel - 2 channel with audible  
SPF22073  
SPF22074  
Sensor for 2050-XX  
Sensor for 2060-XX  
CalibrationAccessories  
1200-26 Gas Sensor Calibrator w/2 gas cylinders  
(specify gas type/conc)  
1290-10 Gas Cylinder - Carbon Monoxide 150 PPM  
1290-11 Gas Cylinder - Hydrogen 80 PPM  
10.0 LIMITED WARRANTY  
In all cases SMC’s responsibility and liability under this  
warranty shall be limited to the cost of the equipment. The  
purchaser must obtain shipping instructions for the prepaid  
return of any item under this warranty provision and com-  
pliance with such instruction shall be a condition of this  
warranty.  
SIERRAMONITOR CORPORATION warrants its products  
to be free from defects in workmanship or material under  
normal use and service for two year after date of shipment.  
SMC will repair or replace without charge any equipment  
found to be defective during the warranty period. Final  
determination of the nature and responsibility for defective  
or damaged equipment will be made by SMC personnel.  
Exceptfortheexpresswarrantystatedabove, SMCdisclaims  
all warranties with regard to the products sold hereunder  
including all implied warranties of merchantability and fit-  
nessandtheexpresswarrantiesstatedhereinareinlieuofall  
obligations or liabilities on the part of SMC for damages  
including but not limited to consequential damages arising  
out of/or in connection with the use or performance of the  
product.  
All warranties hereunder are contingent upon proper use in  
the application for which the product was intended and do  
not cover products which have been modified or repaired  
without SMC approval or which have been subjected to  
accident, improper maintenance, installation or application,  
or on which original identification marks have been re-  
moved or altered. This Limited Warranty also will not apply  
to interconnecting cables or wires, consumables (ie. calibra-  
tion gases, batteries), nor to any damage resulting from bat-  
tery leakage.  
Page:  
11  
2050/2060 Gas Monitors (5/04)  
 

SIIG Switch ID US0011 S1 User Manual
Soleus Air Fan ft1 20 10 User Manual
Sony Cell Phone LT28h User Manual
Sony Cordless Telephone SPP M937 User Manual
Sony MP3 Player TCM 900DV User Manual
Sony VCR SLV 775HF User Manual
Sony Webcam DSC RX100M2 User Manual
Spalding Fitness Equipment M722154 User Manual
StarTechcom Network Cables USB56KEM3 User Manual
Sun Microsystems TV Video Accessories AND B100S User Manual